Geneq > Météorologie > Enregistreurs – Miniatures

Enregistreur de température à 3 canaux
Modèle SmartReader 5


The SmartReader 5 data logger is ideal for those applications that require monitoring extremely high temperatures or monitoring temperature over a wide range. The logger has one internal channel for measuring ambient temperature and two external channels for use with ACR thermocouple wire or any J, K, S or T type thermocouple wire.

Applications
Monitor performance of gas-fired heating equipment, verifying concrete curing temperatures, process verification, etc.

General Specifications
Size 107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”)
Weight 110 g (3.75 oz)
Case Material Noryl® Plastic
Battery 3.6 volt Lithium, 1 Amp-Hour
Resolution 8-bit (1 part in 256)
Mounting Magnetic backing or locking eyelet
Clock Accuracy ± 2 seconds per day
Sampling Methods Continuous (First-in First-out), Stop When Full (Fill-then-stop)
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing)
PC Requirements Windows PC with at least one free USB or serial port (depending on interface)
Software Requirements TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit)
Memory Size 32 KB (capable of storing up to 32,767 readings)
Sampling Rates User selectable rates from 8 seconds to 5 days in 4-second increments
Number of Channels Three (one internal thermistor for cold-junction compensation and two external channels for narrow range and wide range thermocouples)

Sensor Specifications
Internal Temperature Sensor
Type NTC Thermistor - 10,000 Ohm @ 25°C (77°F)
Range -40°C to 70°C (-40°F to 158°F)
Accuracy ±0.2°C over the range of 0°C to 70°C (± 0.3°F over the range of 32°F to 158°F)
Resolution 0.4°C (0.7°F) @ 25°C; better than 1°C (1.8°F) between -25°C and 70°C
(-13°F and 158°F); better than 2.0°C (3.6°F) between -40°C and -25°C (-40°F and -13°F)
Thermocouple Wire Range Resolution
J Type Channel #2 (Wide) -50°C to 600°C (-55°F to 1100°F) 5°C (9°F)
  Channel #1 (Narrow) -20°C to 190°C (0°F to 370°F) 1.4°C (2.5°F)
K Type Channel #2 (Wide) -100°C to 900°C (-145°F to 1650°F) 6.7°C (12°F)
  Channel #1 (Narrow) -25°C to 230°C (-10°F to 440°F) 1.7°C (3°F)
S Type Channel #2 (Wide) 0°C to 1760°C (32°F to 3200°F) 23°C (41.5°F)
  Channel #1 (Narrow) 0°C to 960°C (32°F to 1750°F) 7°C (12.6°F)
T Type Channel #2 (Wide) -200°C to 400°C (-325°F to 750°F) 6.7°C (12°F)
  Channel #1 (Narrow) -35°C to 200°C (-30°F to 390°F) 1.7°C (3°F)
Accuracy ±1% of range + resolution

Ordering Information
Item Cat No.
SR-005 (SmartReader 5) 01-0028