| Food process verification, pharmaceutical storage monitoring, transportation of temperature-sensitive goods,
etc. |
| Size |
107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”) |
| Weight |
110 g (3.75 oz) |
| Case Material |
Noryl® Plastic |
| Battery |
3.6 volt Lithium |
| Resolution |
12-bit (1 part in 4096) |
| Mounting |
Magnetic backing or locking eyelet |
| Clock Accuracy |
± 2 seconds per day |
| Sampling Methods |
Continuous (First-in First-out), Stop When Full (Fill-then-stop)
*Not available with sample rates faster than eight seconds |
| Operating Limits |
-40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing) |
| PC Requirements |
Windows PC with at least one free USB or serial port (depending on interface) |
| Software Requirements |
TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit) |
| Memory Size |
128 KB (capable of storing up to 87,000 readings) |
| Sampling Rates |
User selectable rates from 25 per second to once every eight hours
(BP-101 battery pack or PS-201 power supply required for sampling rates faster
than eight seconds) |
| Number of Channels |
Four (one internal thermistor channel for ambient temperature and three external
channels for use with RTD* temperature probes) *Any RTD with a Temperature
Coefficient of Resistance (TCR) of 0.00385 can be used |
| RTD Channels - The following specifications are for : |
| 4-Wire Platinum 100 Ohm @ 0°C & TCR = 0.00385 |
| 4-Wire Platinum 1000 Ohm @ 0°C & TCR = 0.00385 |
| Temperature |
| Range |
-230°C to 300°C (-382°F to 572°F) |
| Accuracy |
±0.3°C (±0.54°F) + resolution or ±0.06% full scale (of logger) |
| Resolution |
±0.14°C (±0.25°F) |
| Resistance (100 Ohm) |
| Range |
5 to 210 Ohm |
| Accuracy |
±0.13 Ohm, ±0.06% full scale |
| Resolution |
±0.05 Ohm |
| Resistance (1000 Ohm) |
| Range |
50 to 2100 Ohm |
| Accuracy |
±1.3 Ohm, ±0.06% full scale |
| Resolution |
±0.5 Ohm |
| Logger is Temperature compensated by internal temperature sensor for the range of -20°C to 80°C (-4°F to 176°F) |