Geneq > Météorologie > Enregistreurs – Miniatures

Enregistreur de température à 2 canaux
modèle Smart Reader Plus 1


The SmartReader Plus 1 is a powerful and accurate data logger designed for easy monitoring and recording of temperature. With one internal ambient temperature channel and an input for an optional remote temperature or contact closure sensor, the SmartReader Plus 1 is ideal for quick and accurate temperature recordings in a variety of applications.

Applications
Food process verification, pharmaceutical and/or laboratory storage, transportation of temperature-sensitive goods, equipment run time, HVAC system testing and balancing, etc.

General Specifications
Size 107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”)
Weight 110 g (3.75 oz)
Case Material Noryl® Plastic
Battery 3.6 volt Lithium
Resolution 12-bit (1 part in 4096)
Mounting Magnetic backing or locking eyelet
Clock Accuracy ± 2 seconds per day
Sampling Methods Continuous (First-in First-out), Stop When Full (Fill-then-stop)
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing)
PC Requirements Windows PC with at least one free USB or serial port (depending on interface)
Software Requirements TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit)
Memory Size
(3 models available)
32 KB (capable of storing up to 21,500 readings)
128 KB (capable of storing up to 87,000 readings)
1.5 MB (capable of storing up to 1,048,000 readings)
Sampling Rates User selectable rates from 25 per second to once every eight hours (BP-101 battery pack or PS-201 power supply required for sampling rates faster than eight seconds)
Number of Channels Two (One for the internal temperature sensor and one for remote sensing of temperature, resistance, or switch status.)

Sensor Specifications
Internal Temperature Sensor
Type NTC Thermistor - 10,000 Ohm @ 25°C (77°F)
Range -40°C to 70°C (-40°F to 158°F)
Accuracy ±0.2°C over the range of 0°C to 70°C (± 0.3°F over the range of 32°F to 158°F)
Resolution 0.03°C (0.05°F) @ 25°C; better than 0.07°C (0.12°F) between -25°C and 70°C (-13°F and 158°F); better than 0.13°C (0.23°F) between -40°C and -25°C (-40°F and -13°F)
ACR ET Series Thermistor Probes Range Cat. No.
General Purpose Temperature Probe -35°C to 95°C (-30°F to 200°F) 35-0001
Low Temperature Probe -60°C to 55°C (-75°F to 130°F) 35-0028
Low Temperature Penetration Probe -60°C to 55°C (-75°F to 130°F) 35-0000
High Temperature Probe 10°C to 170°C (50°F to 335°F) 35-0002
Oven Temperature Probe 70°C to 255°C (155°F to 490°F) 35-0003
Skin Surface Temperature Probe 20°C to 40°C (68°F to 104°F) 35-0020
Pipe Surface Temperature Probe

-35°C to 95°C (-30°F to 200°F)

35-0025
Waterproof Temperature Probe -40°C to 105°C (-40°F to 221°F) 35-0035

Ordering Information
Item Cat No.
SRP-001 01-0008
SRP-001-1.5 MB 01-0141
SRP-001-128 KB 01-0112