| Performance tracking of pneumatically controlled valves; verifying thermostat and humidistats; monitoring
sensitive shipments, etc. |
| Size |
107mm x 74mm x 22mm (4.2” x 2.9” x 0.9”) |
| Weight |
110 g (3.75 oz) |
| Case Material |
Noryl® Plastic |
| Battery |
3.6 volt Lithium |
| Resolution |
12-bit (1 part in 4096) |
| Mounting |
Magnetic backing or locking eyelet |
| Clock Accuracy |
± 2 seconds per day |
| Sampling Methods |
Continuous (First-in First-out)*, Stop When Full (Fill-then-stop)
*Not available with sample rates faster than eight seconds. |
| Operating Limits |
-40°C to 70°C (-40°F to 158°F) and 0 to 95% RH (non-condensing) |
| PC Requirements |
Windows PC with at least one free USB or serial port (depending on interface) |
| Software Requirements |
TrendReader® 2 (Compatible with Windows XP SP3, Windows Vista SP2& Windows 7 (32 bit & 64 bit) |
Memory Size
(2 memory sizes available) |
32 KB (capable of storing up to 21,500 readings)
128 KB (capable of storing up to 87,000 readings) |
| Sampling Rates |
User selectable rates from 25 per second to once every eight hours
(BP-101 battery pack or PS-201 power supply required for sampling rates faster
than eight seconds) |
| Number of Channels |
Four (one internal sensor for ambient temperature, one for pressure, and two for
optional remote relative humidity and temperature probe [model EH-020A]) |
| Internal Temperature Sensor |
| Type |
NTC Thermistor - 10,000 Ohm @ 25°C (77°F) |
| Range |
-40°C to 70°C (-40°F to 158°F) |
| Accuracy |
±0.2°C over the range of 0°C to 70°C (± 0.3°F over the range of 32°F to 158°F) |
| Resolution |
0.03°C (0.05°F) @ 25°C; better than 0.07°C (0.12°F) between -25°C and 70°C
(-13°F and 158°F); better than 0.13°C (0.23°F) between -40°C and -25°C
(-40°F and -13°F) |
| Pressure Sensor Ranges |
Resolution |
| 0 to 10 inWC (2.5 kPa) Gauge |
0.006 inWC (1.5 Pa) |
| 0 to 5 PSI (30 kPa) Gauge |
0.003 PSI (0.02 kPa) |
| 0 to 30 PSI (200 kPa) Absolute |
0.015 PSI (0.1 kPa) |
| 0 to 30 PSI (200 kPa) Gauge |
0.015 PSI (0.1 kPa) |
| 0 to 100 PSI (700 kPa) Absolute |
0.05 psi (0.34 kPa) |
| 0 to 100 PSI (700 kPa) Gauge |
0.05 psi (0.34 kPa) |
| 0 to 150 PSI (1000 kPa) Gauge |
0.05 psi (0.34 kPa) |
| Calibrated Accuracy |
±0.5% of full scale @ 25°C (77°F) |