GENEQ > Biotechnology > Plate Reader
Microplate Reader
StatFax 2100
| FEATURES | |
|
|
| OPTIONS&ACCESSORIES |
|
| SPECIFICATIONS | |
| Photmetric | |
| Linear Measurement Range | -0.20 to 3.0 Absorbance units (A) |
| Photometric Accuracy | +/- 1% or better |
| Stability | Drift of no more than 0.005A in 8 hours |
| Light Source | Tungsten lamp with lamp saver feature |
| Wavelengths | Standard: 405, 450, 492, and 630nm 6-VIS: 405, 450, 492, 545, 600, and 630nm 6-UV: 340, 405, 450, 492, 545, and 630nm |
| FilterType | Multiple cavity hard coat interference, 10nm half bandpass |
| Vessel | Standard 96-well microtiter plates, or strip trays |
| Electronic | |
| Display | Two line LCD, alpha numeric, 24 characters per line |
| Printer Requirements | Parallel or serial, 80 Column, 2K byte minimun buffer capacity, EPSON compatible |
| Keyboard | 29-key, domed membrane switch, enunciating |
| Power Requirements | 115V or 230V AC, 50-60Hz (switch selectable) |
| Serial Port | 2400 Baud, 1 start bit, 8 data, I stop, no parity |
| Microprocessor | Z80A or Z 180 |
| Non-volatile Memory | Battery supported RAM, holds approx. 50 tests and curves |
| Certifications | NRTL Listed, CE Mark |
| Software | |
| Speed | Reads, and prints absorbances of 96 wells in about two minutes |
| Calculation Modes | Single point calibration, uptake mode, point-to-point curve fit, polynomial regression, linear regression (log & linear), cutoff absorbance, multipoint % absorbance |
| Additional Features | Complete user prompting, flags and error messages, partial plate reading, alphanumeric test naming, auto- matic interpretation options, duplicate options, clock and calendar, controls locator, self check system, curve plotting and editing |
| Other | |
| Mixer | Variable time, fixed speed |
| Enclosure | Fire-retardant ABS plastic cover with metal base |
| Dimensions | Approx. 17xl4.5x7in(43x37xl8cm) weighs 30lbs. (13.7kg) |