Geneq > Essais sur matériaux > Revêtements (Essais sur) > Viscosimètres

Viscosimètre numérique modèle KU-2 Stormer
 


The Digital KU-2 Stormer-Type Viscometer provides a direct digital reading in Krebs units (KU), centipoise and grams (gm). This simplifies an established test procedure of providing a readout in Krebs units, with the added advantage of an immediate calculation of the viscosity value in grams and centipoise.
  • Easy to use – no weights or stroboscopic timing attachment
  • Switch selectable LED digital display of Krebs units or grams or centipoise
  • Allows for rapid and easy measurement of samples
  • Parallel printer output for test documentation
  • Instrument base fits standard pint, 1/2 pint and quart cans
  • Traceable to NIST

The Digital KU-2 Stormer-Type Viscometer can be equipped with an air purge that allows the interior of the instrument to be pressurized with air or inert gas. Recommended when working in a hazardous environment.


Procedure
Samples are tested in paint industry cans that are filled to within 3/4” (20 mm) of the top. The KU-2 Stormer-Type Viscometer operating lever has two positions:
  • Up, where the spindle is stationary and the motor is off
  • Down, when the spindle is rotating and measurement is occurring
Press the Hold switch at any time to freeze the display reading.

Standards
  • ASTM
  • D 562

Specification
Range Resolution Accuracy Repeatability Operating Temp. Spindle Speed
40 - 141 KU; 32 - 1090 grams; 27 - 5274 cps* 0.1 KU; 1.0 gm; 0.7 cps within ± 1 % full scale within ± 0.5 % full scale 10 - 40 °C 200 rpm± 0.1 rpm
40 - 141 KU; 32 - 1090 grams; 27 - 5274 cps* 0.1 KU; 1.0 gm; 0.7 cps within ± 1 % full scale within ± 0.5 % full scale 10 - 40 °C 200 rpm ± 0.1 rpm

Power Supply

115 V / 60 Hz
Dimension 20 x 11 x 15 in
Shipping Weight 20 lbs (10 kg)
*Centipoise values are based on the conversion from Krebs Units as defined in the ASTM standard D562

Comes complete with:
  • Digital KU-2
  • 1 spindle (paddle type)
  • Adapters for 1 pint, 1/2 pint
  • Operation Manual

Accessories
  • Replacement for Digital KU-2 Stormer-Type Viscometer
  • non-reference spindle for relative measurements of pastes