VISTA Abbe Refractometer
Model C10
The VISTA Model C10 Abbe Refractometer is an
excellent, ffordable instrument for the Refractive Index (nD) analysis of
liquids, glass, plastic, & film.![]() |
![]() |
| EASY TO USE |
| Operation consists of placing a sample on the prism; while looking through the eyepiece, the control knob is turned until the shadowline is centered in the crosshairs (see image below); the reading is taken where the vertical line crosses the scale. Calibrates with water. |
| INDUSTRIAL FEATURES |
|
| PRISM TEMPERATURE CONTROL |
| Featuring 2 circulator nozzles for the primary prism and 2 for the secondary prism, model C10 can be connected to a circulating bath to accurately regulate prism temperature. The prism temperature is displayed on the external glass thermometer. The effective temperature range is 0 to 70ºC |
| FILM TESTING |
| Transparent or translucent film samples can be measured with the use of a glass standard (aids in the illumination of the sample), contact liquid (adheres the sample to the glass standard and to the primary prism), and an external sodium light source (sold separately). |
| Note: The refractive index (nD) value of the sample must be lower than that of the contact liquid and the glass standard. The included bromonaphthalene contact liquid has a refractive index (nD) value of 1.6300. The included glass standard has a refractive index value (nD) of 1.5163. |
| GLASS & PLASTIC TESTING |
| Transparent or translucent glass and plastic samples can be measured with the use of a contact liquid (adheres the sample to the primary prism), and an external sodium light source (sold separately). |
| Note: The refractive index (nD) of the sample must be lower than that of the contact liquid. The included bromonaphthalene contact liquid has a refractive index (nD) value of 1.6300. |
| SPECIFICATIONS | |
| Measuring System | Optical refraction critical angle detection |
| Measuring Range | Refractive index: 1.3000-1.7000 Brix: 0.0-95.0% |
| Resolution | Refractive index: 0.0005 Brix: 0.25% |
| Accuracy | Refractive index: ±0.0003 Brix: 0.15% |
| Dispersion (nF-nC) | Dial Readings w/conversion charts |
| Ambient Temperature | 5-35ºC |
| Measuring Temperature | 0.0-70.0ºC |
| Calibration Liquid/Medium | Distilled water; glass standard, 1.5163 nD (w/bromonaphthalene contact liquid) |
| Prism | Optical glass |
| Prism Housing | SUS316 stainless steel |
| Circulator Nozzles | 2 for primary prism, 2 for secondary prism |
| Dimensions (LxWxH) | 14 x 9 x 24 cm |
| Weight | 2.3kg |