VelociCalc Multi-Function Thermal Anemometer
Model 9555


Model 9555
Features and Benefits
• Best in class accuracy, especially in low flows
• Displays up to 5 measurements simultaneously
• Optional “smart” plug-in probes, including CO2 and rotating vane probes
• Large graphic display
• Manual or continuous data logging
• TRAKPROTM downloading software included
• Name test IDs meaningful to you
• Bluetooth® printer capability
• Fast calibration and repair service—just send in the probe

The Model 9555 series are portable, hand held, Multi-Function Ventilation Test Instruments. These instruments are available with or without a differential pressure sensor and are designed to work with a wide range of plug-in probes. The probes allow users to make various measurements by simply plugging in a different probe that has the features and functions best suited for a particular application. They are designed to measure air velocity, temperature, humidity, CO and CO2. Calculations include air flow, heat flow, turbulence, wet bulb and dew point temperature. The probes can be ordered at any time and include a data sheet with certificate of traceability. When its time for servicing, only the probe needs to be returned since all the calibration data is stored within the probe.

Applications

• HVAC commissioning and troubleshooting
• Clean room certification
• Testing and balancing
• Ventilation evaluations
• Thermal comfort studies
• IAQ investigations
• Process air flow testing

Specifications
VELOCICALC
Models 9555, 9555-A, 9555-P, and 9555-X
Velocity (TA Probe)
Range 0 to 9,999 ft/min (0 to 50 m/s)
Accuracy1&2 ±3% of reading or ±3 ft/min (±0.015 m/s), whichever is greater
Resolution 1 ft/min (0.01 m/s)
Velocity (Pitot Tube for Meter Models 9555, 9555-A, 9555-P)
Range3 250 to 15,500 ft/min (1.27 to 78.7 m/s)
Accuracy4 2,000 ft/min (±1.5% at 10.16 m/s)
Resolution 1 ft/min (0.01 m/s)
Velocity (Rotating Vane Probe Model 995)
Range 50 to 6,000 ft/min (0.25 to 30 m/s)
Accuracy ±1% of reading or ±4 ft/min (±0.02 m/s)
Resolution 1 ft/min (0.01 m/s)
Duct Size
Dimensions 1 to 250 inches in increments of 0.1 in.
(1 to 635 cm in increments of 0.1 cm)
Volumetric Flow Rate
Range Actual range is a function of velocity, pressure, duct size, and K factor)
Temperature (TA Probe Models 964, 966, 980, 983)
Range 14 to 140°F (-10 to 60°C
Accuracy5 ±0.5°F (±0.3°C)
Resolution 0.1°F (0.1°C)
Temperature (TA Probe Models 960, 962)
Range 0 to 200°F (-18 to 93°C)
Accuracy5 ±0.5°F (±0.3°C)
Resolution 0.1°F (0.1°C)
Temperature (Rotating Vane Probe Model 995)
Range 32 to 140°F (-0 to 60°C)
Accuracy5 ±2.0°F (±0.1°C)
Resolution 1.0°F (0.1°C)
Relative Humidity (TA Probe Models 964, 966, 980, 982)
Range 32 to 140°F (-0 to 60°C)
Accuracy6 ±2.0°F (±0.1°C
Resolution 1.0°F (0.1°C)
Relative Humidity (TA Probe Models 964, 966, 980, 982)
Range 0 to 95% RH
Accuracy6 ±3% RH
Resolution 0.1% RH
Carbon Monoxide (IAQ Probe with CO Model 982)
Range 0 to 5000 ppm
Accuracy6 ±3% of reading or ±50 ppm, whichever is greater
Resolution 1 ppm
Static/Differential Pressure (Probe for Meter Models 9555, 9555-A, 9555-P)
Range7 -15 to +15 in. H2O (-28.0 to +28.0 mm Hg,
-3735 to +3735 Pa)
Accuracy ±0.01 mm Hg, ±0.005 in. H2O (±1% of reading ±1 Pa)
Resolution 0.001 in. H2O (1 Pa, 0.01 mm Hg)
Barometric Pressure
Range 2.036 to 36.648 in. Hg
(517.15 to 930.87 mm Hg)
Accuracy 2% of reading
Instrument Temperature Range
Operating (Electronics) 40 to 113°F (5 to 45°C)
Storage -4 to 140°F (-20 to 60°C)
Data Storage Capabilities
Range 26,500+ samples and 100 test IDs
Logging Interval
1 second to 1 hour 
Time Constant
User selectable 
External Meter Dimensions 
3.8 in. x 8.3 in. x 2.1 in. (9.7 cm x 21.1 cm x 5.3 cm) 
Meter Weight with Batteries
0.8 lbs. (0.36 kg)
Meter Probe Dimensions
Probe Length 40 in. (101.6 cm)
Probe Diameter of Tip 0.28 in. (7.0 mm)
Probe Diameter of Base 0.51 in. (13.0 mm)
Articulating Probe Dimensions
Articulating Section Length 6.4 in. (16.26 cm)
Diameter of Articulating Knuckle 0.38 in. (9.5 mm)
Power Requirements Four AA-size batteries or AC adapter

Optional Probes for VELOCICALC 9555 Series Multi-Functional Anemometers
Model
Probe Description
960
Air Velocity and Temperature, straight probe
962
Air Velocity and Temperature, articulating probe
964
Air Velocity, Temperature, and Humidity, straight probe
966
Air Velocity, Temperature, and Humidity, articulating probe
995
100 mm Rotating Vane probe
968
Draft and Comfort probe
972
Surface Temperature probe
974
Air Temperature probe
980
Indoor Air Quality probe
982
Indoor Air Quality probe, with CO

 
9555
9555-A
9555-P
9555-X
Probe that measures velocity, temperature, and humidity
straight
articulated
optional
optional
Pressure measurement
Calculates flow, wet bulb, dew point, standard/actual
optional
optional
Optional velocity and temperature probe
Optional rotating vane probe
Optional IAQ probes (CO2,
temperature, humidity, CO)
Data Logging (manual, auto save continuous)
Data logging software
Free Certificate of Calibration
• 1 Temperature compensated over an air temperature range of 5 to 65°C (40 to 150°F).
• 2 The accuracy statement begins at 0.15 m/s through 50 m/s (30 ft/min through 9,999 ft/min).
• 3 Pressure velocity measurements are not recommended below 5 m/s (1000 ft/min) and are best suited to velocities over 2,000 ft/min (10.00 m/s). Range can vary depending on barometric pressure.
• 4 Accuracy is a function of converting pressure to velocity. Conversion accuracy improves when actual pressure values increase.
• 5 Accuracy with instrument case at 25°C (77°F), add uncertainty of 0.03°C/°C (0.05°F/°F) for change in instrument temperature.
• 6 Accuracy with probe at 25°C (77°F). Add uncertainty of 0.2% RH/°C (0.1% RH/°F) for change in probe temperature. Includes 1% hysteresis.
• 7 Overpressure range = 190 in. H2O, 360 mmHg, 48 kPa.