GENEQ > Biotechnologie > Activité de l’eau - Mesure

Appareil de mesure de l'activité de l'eau
modèle ms1 Set aw de Novasina

ms1 Set aw
ms1 Set AMS
Accessories for Set aw/M/AM/AMS
Replacement parts for ms1 Set aw

Options for instrument ms1

A hand-held precision instrument for the measurement of aw values
(water activity) in foodstuffs and pharmaceutical products.

A handy, easy-to-use, self-powered water activity measurement instrument set. Functions, test and precision can be checked at all times with saturated salt solution standards. The instrument is supplied in a practical plastic case, together with all required accessories.


aw range:
0.06 ... 0.98aw, not condensing
Temperature:
-20...+80°C
Sensor:
enBSK-3 with a resistive electrolyte measurement cell
Precision:
+/- 0.01aw, +/- 0.3°C
Communication:
RS232 interface (optional)

ms1 Set aw

Precise, fast, flexible

A practical, 1'000-times proven precision hand-held measurement instrument for aw / relative humidity and temperature measurements. Thanks to the Novasina electrolyte-Sensor it is well-suited for precise aw measurements in many types of products. The instrument and its sensor are very robust, have excellent long-term stability and are practically hysteresis-free.

The large and clear LC display is very easy to read. The instrument is powered by a 9 V battery powered, but an additional mains power supply is available to support long-term measurements. The sealed measurement probe cell is not temperature-controlled. The measurement temperature can be displayed at any time on the large LCD display. The instrument has an indicated stability value display function. An optional RS 232 interface and supporting software is available for connecting a printer or PC.

The instrument is delivered in a sturdy plastic carrying case with space for various additional accessories. Measurements can be made with the instrument in its case.

Examples of aw measurements with the Set aw:

  • All forms of baked goods
  • Meat, sausages and cheese
  • Dried foodstuffs, fruit concentrates
  • Cosmetics
  • Pharmaceuticals, natural medicines
Instrument ms1:
Dimensions : 54x43x265mm
Weight : 280 g
Power : 9V Battery
Humidity sensor:
Electrolyte-Measurement cell
Range : 0.06...0.98aw
Repeatability : +/- 0.003aw
Precision :+/- 0.01aw
(0.10...0.95aw, 25°C, with 5 point calibration)
Temperature sensor:
Precise NTC-Resistance
Range : -20...80°C
Repeatability :+/- 0.1° C
Precision : +/- 0.3° C
Humidity standards:
Salt tablets SAL-T:
Saturated pure salt solutions, based on national standards are recommended, with the following humidity values:
11%, 33%, 53%, 75%, and 90% rh

Consisting of:

111 0137 ms1 Set AW 230V 111 0150 ms1 Set AW 115V

The Set aw is a simple and precise entry-level set for important aw value measurements in the food sector. It consists of an ms1 hand-held instrument with a sealed measurement chamber, and a range of accessories all housed in a robust plastic case. The latter also has space for the optional RS232 interface for attachment to a PC. SAL-T humidity standards are included for testing and recalibrating the measurement cell as required.

Carrying case with:

  • Measurement instrument ms1 aw
  • Sensor enBSK-3/PP, with 0.5m cable
  • 9V Battery
  • 5 SAL-T: 11, 33, 53, 75, 90% rh
  • 40 disposable sample dishes ePW
  • Tweezers
  • User manual
  • Manufacturers report for the 5-point calibration 11-33-53-75-90%
  • Weight: 2.900 kgs

111 6574 ms1 Set AMS

The Set AM is intended primarily for aw measurements in very large volumes of foodstuffs such as cereals, coffee beans or cheese, or pharmaceutical powders etc. A 13 mm SARK insertion probe is included for this purpose The system is supplied in a robust plastic case including a range of accessories. Space is available for further options, such as a power supply or interface/cable. The Set can be tested and recalibrated as required using individual Sensor Check humidity standards.

Carrying case with:

  • Measurement instrument ms1 aw
  • Sensor enSARK-3, 2m cable
  • Sharp sensor tip
  • 9V battery
  • Plastic rack for the ms1
  • Sensor check SC 75% rh, with adapter ch
  • Polystyrene box for calibration
  • User manual
  • Manufacturers report for the 5-point calibration 11-33-53-75-90%
  • Weight: 2.300 kgs

Accessories for Set aw/M/AM/AMS

111 0826 Interface PC

PC communication via RS232
RS232 interface, 1m long cable, CD with software NOVALOG32 for Win 9x/2000/NT/XP.

Interface Set comprising:

  • 1m cable with D-Sub 25/25-pin
  • RS-232 Interface
  • CD with NOVALOG32 software for Windows systems
  • User manual
  • Weight: 450g

111 0562 Interface Seiko

RS232 interface for Seiko printer

RS-232 interface, 1m long cable to connect a Seiko DPU 411 thermal printer. This makes it possible to control the printer using the printer parameters defined in the ms1 menu.

The printer must be obtained locally

Printer-Interface-Set Seiko:

  • Printer cable : Flat cable
  • 1m long printer cable with D-Sub 25/25-pin
  • RS 232 interface
  • Seiko adapter box
  • Weight : 300g
 

111 6256 Adapter 25F/9M

D-Sub adapter 25F/9M
Additional adapter for the RS232 Interface for Seiko printer (111 0562), permitting connection to the more recent Seiko DPU 414 printer.

Adapter Interface:

  • Dimensions : 60 x 40 x 20 mm
  • Adapter D-Sub 25F/9M
  • 25-pin female con.
  • 9 - pin male con.
  • Weight: 50g
 

111 0661 Interface Epson

RS232 interface for Epson printer

RS-232 interface, 1m long cable to connect an Epson or other compatible 9-pin needle printer. This makes it possible to control the printer using the printer parameters defined in the ms1 menu.

The printer must be obtained locally.

Printer-Interface-Set Epson:

  • Printer cable: Flat cable
  • 1m long printer cable with D-Sub 25/25-pin
  • RS 232 interface
  • Epson adapter
  • Weight: 300g
 

Example - Seiko Printer

Local printer system:c

Any printer must be obtained locally.

Novasina does not supply printers.

Printers:

Portable printer for an easy direct connection to an ms1 hand-held instrument via a serial interface. A range of different printers can be supported by the ms1.

 

111 8426
Seiko Thermal paper
Thermal paper replacement rolls

for Seiko DPU 411, 414 printers.

Attention: This paper must be stored in a dark and dry place. This is thermal transfer paper only intended for Seiko printers.

Printer type: Seiko DPU 411, 414

Note: Set of 5 rolls, each 25m long

Thermal paper replacement:

Set of 5 rolls, each 25 m long. Thermal paper for Seiko printer Models DPU 411 and 414

  • Paper width 11cm
  • Weight: 850g each Set
  • Set of 5 rolls, each 25m long
 

111 0601 ePW sample dishes

Sample dishes ePW

For an enBSK measurement chamber

Packet of 100 sample dishes for the sample sensor enBSK-3.

These are disposal, intended for a single use with any Novasina aw instrument

A package consists of:

  • 100 sample dishes ePW
  • Material : Polypropylene
  • Dimensions : 40mm diameter x 12mm deep
  • Weight : 440g
 

111 0790 SI Set BS

SI, sorption isotherm set for the BSK Sensor

This accessory, together with the Novasina humidity standards SAL-T, makes it possible to easily determine points of a sorption isotherm. (A precision balance and an oven are also required.)

Prerequisite: the laboratory room temperature must be kept absolutely constant.

SI – Set BS complete with:

  • 1 SI-ring BS
  • 1 substance basket
  • 1 weighing glass
  • 1 pair of tongs For precise determination of a sorption isotherm, the AW SPRINT or AW CENTER with temperature control should be used.
  • Weight: 280g
 

111 1218 Tweezers for CK cells

Special tweezers
To simplify the exchange of a CK measuring cell in an enBSK measurement chamber.

Special-tweezer:

  • Material : stainless steel
  • For changing the humidity cell in an enBSK chamber
  • Weight: 20g
 

111 6453 SAL-T 6% rh
111 0930 -> SAL-T 11% rh
111 0932 -> SAL-T 33% rh
111 0934 -> SAL-T 53% rh
111 0936 -> SAL-T 75% rh
111 0938 -> SAL-T 90% rh
111 0940 -> SAL-T 98% rh

SAL-T salt tablets:

Humidity standards based on saturated salt solutions in plastic tablets with humidity-permeable membranes. Each salt is delivered in an airtight can. SAL-T are available with the following humidity values (at 25°C):

06,4 % Colour white
11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour purple
90.1 % Colour white
98.0 % Colour orange

Humidity values in the temperature range
15° .... 30°C:

06.9 ..... 06.2% rh
11.3 ..... 11.3% rh
33.3 ..... 32.4% rh
55.9 ..... 51.4% rh
75.6 ..... 75.1% rh
90.9 ..... 89.9% rh
98.5 …. 98.0% rh

The precision corresponds to the Greenspan Report 1977
typically +/- 0.3 % rh

  • Weight: 50g a piece
 

111 7000 -> SAL-T/C 6% rh
111 1056 -> SAL-T/C 11% rh
111 1057 -> SAL-T/C 33% rh
111 1060 -> SAL-T/C 53% rh
111 1063 -> SAL-T/C 75% rh
111 1065 -> SAL-T/C 90% rh
111 1068 -> SAL-T/C 98% rh
Salt tablets SAL-T/C EU with European certificate

Humidity standards based on saturated salt solutions in plastic tablets with humidity-permeable membranes. Each salt is delivered in an airtight can. SAL-T are available with the following humidity values (at 25°C):

06,4 % Colour white
11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour purple
90.1 % Colour white
98.0 % Colour orange

Internationally accredited laboratory

  • All Novasina humidity standards are also available with an internationally recognised certificate from an accredited European laboratory (UKAS England).
  • Weight: 50g a piece
 

111 0695 Case with 5 SAL-T

Set of 5 salt tablets SAL-T

Humidity standards based on saturated salt solutions in plastic tablets with humidity-permeable membranes. Each salt is delivered in an airtight can. The set consists of a case containing one each of the following values (at 25°C):

11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour violet
90.1 % Colour white

Case with 5 salt tablets:
SAL-T for 11 ... 90%rh
Humidity values in the temperature range 15° .... 30°C:

11.3 ..... 11.3% rh
33.3 ..... 32.4% rh
55.9 ..... 51.4% rh
75.6 ..... 75.1% rh
90.9 ..... 89.9% rh

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

  • Weight: 670g
 

111 0435 Case with 6 SAL-T

Set of 6 salt tablets SAL-T

Humidity standards based on saturated salt solutions in plastic tablets with humidity-permeable membranes. Each salt is delivered in an airtight can. The set consists of a case containing one each of the following values (at 25°C):

11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour violet
90.1 % Colour white
98.0 % Colour orange

Case with 6 salt tablets
SAL-T for 11 ... 90%rh
Humidity values in the temperature range 15° .... 30°C:

11.3 ..... 11.3% rh
33.3 ..... 32.4% rh
55.9 ..... 51.4% rh
75.6 ..... 75.1% rh
90.9 ..... 89.9% rh
98.5 ….. 98.0% rh

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

  • Weight: 710g
 

Certification of an aw measurement instrument in an accredited European laboratory
A UKAS-laboratory will certify the instrument at two or more humidity values and at several different temperatures if required.

Internationally accredited laboratory

Certified instruments can be supplied!!

 

111 1001 -> eVC-21
111 1003 -> eVC-26
111 7212 -> Redox
111 0995 -> eVALC
111 0997 -> eVMT-2
111 0999 -> eVMS-2
Cell protection filters
A number of filters are available to protect the measurement cell from abrasion or corrosive vapors.

Filters for enBSK head:

111 1001 -> eVC-21 111 1003 -> eVC-26 111 7212 -> Redox 111 0995 -> eVALC 111 0997 -> eVMT-2 111 0999 -> eVMS-2

  • Weight: 30g
Replacement parts for ms1 Set aw

111 1083 9V battery

9V-battery

Alkali-manganese battery (not rechargeable)

This battery permits up to 25 hours of continuous operation.

If a rechargeable battery is used, an external charger is required. The ms1 cannot perform this function.

9V Battery for the ms1:

  • Alkali cell battery :
  • Battery 9V DC, 180mAh
  • A standard battery, available in most electronic shops
  • Max. storage time : 2 years
  • Weight: 50g
 

111 1564 230V power supply 111 1406 115V power supply

230V or 115V power supply

The power supply is recommended when the ms1 is used for a long period in one place, or to preserve battery power when the RS 232 interface is extensively used.

The ms1 does not recharge batteries. If rechargeable batteries are used an external charger is needed.

Power supply:

  • 230V AC +/- 10% / 50-60Hz
  • 115V AC +/- 5% / 50-60Hz
  • Output : 9 V DC / 16 mA
  • Power requirement: 0.4 VA
  • Plug : 4 mm diameter coax
  • Weight: 250g
 

111 1013 CK-3 cell

Replacement measurement cell CK-3

For the enSARK-3, enCSK-3 sensors

Humidity measurement cell with neither temperature measurement nor integrated saturation protection. The aluminium (SARK) or sinter (CSK) filter should be removed and the measurement cell replaced by hand. In no circumstances should, the filter surface on the top of the cell be touched by hand or a hard object.

Humidity measurement cell. The ms1 requires recalibration after cell replacement.

Humidity-Measurement cell CK-3:

  • Humidity measurement
  • Range : 6.....95 % rh
  • Attention : No saturation protection!
  • Repeatability: 0.3 % rh
  • Precision : +/- 1.0 % rh (when calibrated on instrument on all 5-Points from 11 to 90%rh)
  • Weight: 1g
 

110 6821 enBSK-3

Sample sensor enBSK-3
Ideal for small sample volumes. For use with the measuring plate PP and the disposable sample dish ePW. The sensor consists of the humidity measurement cell CK-3, the NTC-temperature sensor and a braided protection filter.

Sample sensor enBSK-3:

  • Material : Aluminum anodised
  • Weight : 240g
  • Connector : 9 pins
 

111 1373 O-Ring : 1
111 2934 Braided filter : 2
111 0111 Tension ring : 3

Replacement parts for measurement head BSK
1 O-ring seal for the measurement chamber.
Recommendation: The O-ring should be periodically exchanged and lubricated with silicon grease.
2 Braided filter for protection from particles.
3 Tension ring for clamping the filter(s).

Spare parts to enBSK:

  • O-ring seal
  • Mesh filter
  • Tension ring
  • Weight

    1 — 1

    2 — 30g

    3 — 10g

 

110 0770 Measurement plate

Sample measuring plate PP for enBSK-3
The product sample is placed in a sample dish ePW (111 0601), which is inserted in the sample measurement plate.

Measurement plate PP: for enBSK sensor head

  • Dimensionens : 15 x 9 mm
  • Material : Anodised aluminium
  • Weight : 200g
 

110 6515 Tweezers for ePW

Tweezers for ePW
To ease the removal of the sample dish ePW from the sample measurement plate PP.

Tweezers:

  • Material : stainless steel
  • Weight: 20g
 

110 6562 Polycarbonate tubes

Set of 10 external tubes For use with enSARK-3 sensor
Set of 10 precision, reusable tubes with closing caps for humidity measurements in concrete floors or walls using the enSARK-3 sensor. The tubes are inserted in boreholes in the concrete floors or walls. The humidity is measured with a SARK sensor once a stable value has been reached.

Set of 10 Precision tubes:
incl. closing caps:

Application:

  • Make an opening in the wall/floor.
  • Insert the tube.
  • Wait for the humidity to stabilise.
  • Measure the humidity using an ms1 and a SARK Sensor (See also Set AM )
  • Weight: 20g
Options for instrument ms1

111 6716 enSARK

Replacement sensor enSARK-3
with 2m connection cable and 9 pin plug

A 13 mm diameter sensor rod with a Duroplast tip, CK-3 humidity measurement cell, and NTC temperature sensor. Designed for measurements in, for example, material in granulated form in sacks, tanks and silos.

See also Set AM, Set AMS:

Example: humidity measurements of granulates, sawdust, or tobacco in a sack

  • Weight: 100g

110 7345 Adapter CH

Sensor-Check adapter ch
for any 13mm sensor to SC-check salt standard

This adapter is needed for calibration. It ensures an airtight connection of a Sensor Check to the sensor tube of the top of the instrument. It reduces the diameter of 20 mm to 13 mm and gives a tight seal.

The adapter should first be attached to the sensor and then to the SC check. Both seals should be carefully checked, leaks will bias the calibration.

CH adapter for any 13mm sensors to SC checks:

  • Dimensions : 30 x 13 mm
  • Material : Polycarbonate
  • Weight : 5 g
  • Quantity : 1 unit

111 6497 enSARK tip

Replacement tip for Sensor enSARK-3

Replacement part for the SARK sensor for humidity measurements in granulates, grains, paper etc. A hard sharp plastic tip to ease the insertion of SARK sensors into materials

Cannot be used with older SARK sensors

Sharp tip for SARK sensor:

  • Material : POM black
  • Weight: 3g

111 0852 Conical sensor

Conical sensor enCSK-3

Conical sensor for measurement in vats, or directly in materials. Incorporates a sintered filter to protect the measurement cell.

Range: 6..95% rh / -20....+80°C.

With a 1 meter cable and 9-pin plug

Conical sensor to ms1:

  • Cable length : 1m cable
  • Weight: : 270g
  • Connector : 9-pin plug to the ms1 instrument
  • Meas. cell : CK-3 with NTC for temp. measurement
  • Range : 6..95%rh/-20..+80°C
  • Can also be built into a box sensor, see also ms1 set M

111 0526 Box sensor enMBRK-3

Box sensor enMBRK-3

A special sensor for relative humidity/aw measurements of larger quantities of foodstuffs, e.g. potato chips, tea leaves, granulates.

Includes the conical sensor enCSK-3 (Nr. 111 0852)

Box sensor enMBRK-3 (incl. Conical sensor enCSK-3)

  • Cable length : 1m
  • Weight : 270g
  • Connector : 9-Pin male for ms1

See also Set M

111 1467 Sintered filter

Sintered filter for conical sensor
For the conical sensor enCSK-3 (111 0852) and box sensor (111 0526). Attention: When changing the filter, the sensor must be treated with extreme care!

Sintered filter for conical sensor:

  • Dimensions : 15mm x 25mm
  • Material : Sintered brass
  • Weight : 30 g

111 1023 adapter cs

Adapter cs for enCSK-3 sensor to SC-check

This adapter is needed for calibration of the enCSK sensor with the salt SC checks. It ensures an airtight connection of a Sensor Check to the conical sensor.

Adapter CS only for the conical sensor Attach the adapter to the conical sensor first and then to the SCcheck.Check that it is well-sealed.

Adapter cs to conical sensor:

  • Dimensions : 30 x 13 mm
  • Material : PE plastic
  • Weight : 5 g

The adapter has no gasket on the sensor side The cone itself provides an airtight seal.

111 0885 -> 11% rh
111 0855 -> 33% rh
111 0857 -> 53% rh
111 0859 -> 75% rh
111 0896 -> 90% rh

Sensor-Checks SC:

Humidity standards based on saturated salt solutions in plastic cylinders with moisture permeable membranes. Each salt is delivered in a well-sealed box. Sensor Checks SC are obtainable for the following values (at 25°C):

11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour purple
90.1 % Colour white

Humidity values in the temperature range 15° .... 30°C:

11.3 ..... 11.3% rh
33.3 ..... 32.4% rh
55.9 ..... 51.4% rh
75.6 ..... 75.1% rh
90.9 ..... 89.9% rh

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

  • Weight: 90g a piece

111 1044 -> 11% rh
111 1037 -> 33% rh
111 1040 -> 53% rh
111 1035 -> 75% rh
111 1032 -> 90% rh

Sensor-Checks SC with European certificate:

Humidity standards based on saturated salt solutions in plastic cylinders with moisture permeable membranes. Each salt is delivered in a well-sealed box. Sensor Checks SC are obtainable for the following values (at 25°C):

11,3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour purple
90.1 % Colour white

Internationally accredited laborator:

All Novasina humidity standards can also be supplied with an internationally recognised certificate from an accredited European laboratory (UKAS England).

Weight: 90 g a piece

111 7847 Sensor check set

Set with 5 Sensor-Checks SC:

Humidity standards based on saturated salt solutions in plastic cylinders with moisture permeable membranes. Each salt is delivered in a well-sealed box. Sensor Checks SC are obtainable for the following values (at 25°C):

11.3 % Colour white
32.8 % Colour blue
52.9 % Colour green
75.3 % Colour purple
90.1 % Colour white

Case with all 5 Sensor Checks from 11 to 90%rh:
Humidity values in the temperature range 15° .... 30°C:

11.3 ..... 11.3% rh
33.3 ..... 32.4% rh
55.9 ..... 51.4% rh
75.6 ..... 75.1% rh
90.9 ..... 89.9% rh

The precision corresponds to the Greenspan Report 1977 typically +/- 0.3 % rh

  • Weight: 700g

111 3542 Extension cable

Extension cable for ms-1

1.5m long cable with 9-pin connectors for remote measurements with the ms1.

Use of the extension cable does not require recalibration of the instrument

Extension cable to ms1:
  • Extension cable: 1.5m
  • Weight: 120g
    Connector: 9-pin special connectors for ms1and sensor.

111 0898 Surface temp. sensor

Surface temperature sensor
Separate (NTC) temperature sensor for measuring surface temperatures of walls etc. This plugs into the side of the ms1 and disables the built-in temperature sensor of the ms1.

Surface temperature sensor:

  • Range : -20...80°C
  • Repeatability : +/- 0.1° C
  • Precision : +/- 0.3° C
  • Weight : 80g

111 0846 HOG-00 Extension

Extension tube HOG- 00

Fits the ms1 basic instrument and enCK-3 sensor. Connects the sensor to the ms1 for applications in air ducts and enclosures.

Use of the extension tube does not require recalibration of the instrument.

Extension tube HOG-00:

  • Length : 240 mm
  • Weight : 80g
  • Material : Aluminium / Polycarbon

111 0978 enCK-3

Sensor enCK-3

Sensor with resistive electrolyte humidity measurement cell CK-3 and integrated temperature probe. Standard humidity sensor for the ms1.

A precise humidity and temperature sensor for 6..95 % rh (98% non-condensing) without internal saturation protection. The integrated, precise NTC temperature probe has a range of –20 to +80 °C. The humidity sensor is protected by an integrated chemical filter.

Information: recalibration of the ms1 at a minimum of two points is required when a sensor is replaced. High precision requires a 5-point calibration.

Sensor enCK-3:

  • Dimensions : 58 x 22mm
  • Weight : 10 g
  • Humidity measurement:
  • Range: 6.....95 % rh Attention: No saturation protection!
  • Repeatability: 0.3 % rh
  • Precision: +/- 1.0 % rh when calibrated on instrument on all 5-Points from 11 to 90%rh.
  • Temperature Measurement: Range temp : -20...80°C Repeatability : +/- 0.1° C Precision : +/- 0.3° C
top of the page