Shopping Cart

0 item(s) - 0,00$
Your shopping cart is empty!
 
Drop us a line
Contact
Handheld Radiation Alert® Detector

Handheld Radiation Alert® Detector

Product Code:DIGILERT200


Need more Info? (514) 354-2511 - 1-800-463-4363

The Digilert 200 measures alpha, beta, gamma, and x-rays. Its digital display shows readings in your choice of counts per minute (CPM), CPS, μSv/hr, mR/hr, or in accumulated counts. Now with USB and the included Observer USB Software, you can download your accumulated data from the Digilert200’s internal memory, set alarms, and calibrate your instrument!
 

Detector
Halogen-quenched GM tube with thin mica end window. Mica window density 1.5-2.0 mg/cm2. Effective window diameter is .360 inch. Side wall is .012 inch thick.

 

Operating Range
  • mR/hr - .001 (1μR) to 200 mR/hr
  • μSv/hr - .01 to 2000
  • CPM - 0 to 214,000
  • CPS - 0 to 3575
  • Total/ Timer - 1 to 9,999,000 counts

 

Accuracy
±10% typical (NIST), Typically ±15% of reading.

 

Energy Sensitivity
1070 CPM/mR/hr referenced to Cs137
 
  • Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%.
  • Detects beta at 50 keV with typical 35% detection efficiency.
  • Detects beta at 150 keV with typical 75% detection efficiency.
  • Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the case.

 

Specifications
Display Backlit 4-digit liquid crystal display with mode indicators.
Display updates every 3 seconds.
Count Light Red LED flashes with each count.
Audio Indicator Sounds with each count (can be switched off for silent operation)
Alarm Pulsating beeper sounds the alarm. Three separate adjustable alert levels are used for mR/hr, CPM, μSv/hr, and CPS.
Outputs Dual miniature jack sends counts to CMOScompatible devices, including computers, data loggers, and educational data collection systems. USB for use with Observer USB Software for PCs
Anti-Saturation Readout will hold at full scale in radiation fields as high as 100 times the maximum reading.
Power Requirements One 9-volt alkaline battery. Battery life is approx.
2000 hours at normal background radiation levels.
Temperature Range -10° to +50°C (14° to 122°F)
Weight 217 g (7.7 oz.)
Size 150 x 80 x 31 mm (6.1 x 3.1 x 1.2 in.)
Options Xtreme Boot & Stand
Limited Warranty 1 year limited warranty

 

Digilert 200 shown in Xtreme Boot Above The Xtreme Boot option
provides easy access to the buttons.

USB Observer Software collects data stored in the internal memory of The InspectorUSB

 

Utility Menu
The Utility Menu offers several options accessible by holding down the + button while turning the instrument on.
 
  • 3 second (fast response)
  • Switches from mR/hr and CPM to μSv/hr and CPS
  • Reset the cal factor to 100.
  • Adjust the CAL factor.
  • Reset all setting to default factory setting.

 