 Analyzer for oils and fats Material Testing

Analyzer for oils and fats

Product Code:InfraCal TPH/TOG


Need more Info? (514) 354-2511 - 1-800-463-4363

Features
  • Measurements in under 10—15 minutes, including extraction 
  • Detect levels as low as 2 ppm or as high as 5,000+ ppm (measurement range dependant on the InfraCal Model selected) 
  • Measures both aliphatic and aromatic components in the sample
  • Models usable with hexane, pentane, Vertrel MCA, perchloroethylene, Freon-113, AK-225 o r S-316 extraction procedures 
  • Ideal for field or laboratory applications 
  • Direct reading, reproducible measurements 
  • Portable, rugged, lightweight, easy-to-use

 

Whether your application is
  • analyzing produced water on offshore or onshore oil rigs 
  • monitoring refinery or industrial plant wastewater effluents 
  • measuring FOG discharge levels into treatment centers 
  • determining the efficiency of oil/water separation systems 
  • soil studies at remediation sites or around underground storage tanks 
  • measurement of residual oil on precleaned metal components 
  • determining the purity level of reclaimed solvents 
  • or virtually any on-site testing of water and soil to determine fat/oil/grease concentration levels

 

InfraCal TOG/TPH analyzers provide an easy infrared analysis method for on-site measurement of total oil and grease (TOG), and total petroleum hydrocarbon (TPH) in water and soil samples and fat, oil and grease (FOG) levels in water samples. Depending on sensitivity requirements and the extracting solvent being used, several InfraCal models are available for TOG/TPH analysis - horizontal platform (HATR-T2), cuvette holder (CVH) and card reader (CH). All models are portable (operable from a 12 volt dc power supply), weigh less than 5 lbs, and read out directly in concentration level. Their compact, rugged design has no moving parts requiring alignment or adjustment. And, user friendly operation makes them easy to use with minimal training. InfraCal TOG/TPH Analyzers can read out infrared absorption values or they can be calibrated to display concentrations directly in any desired format such as ppm or milligram/kg.

A range of optional accessories expands the capabilities and versatility of the InfraCal TOG/TPH Analyzers. The battery pack that provides up to 16 hours of continuous operation enables the analyzer to be used for field testing. A small, portable printer further enhances the analyzer’s capability for on-site field use providing sequenced and time stamped printouts of analytical results. In addition, the calibration table can be printed giving a complete record for a series of analyses. Transporting the InfraCal TOG/TPH Analyzer and various accessories is easy with our sturdy carrying case with pre-diced pluck foam inserts. 

 

InfraCal IR Platform Analyzer, Model HATR-T2 (P/N 405-1009)
This model is recommended for measuring TOG, TPH or FOG concentration levels when using hexane, pentane or Vertrel MCA as the extracting solvent. Measurement data obtained with this analyzer will correlate to EPA Method 1664 as both procedures are based on evaporation techniques and measuring the residual oil and grease. The analyzer is equipped with a built-in cubic zirconia horizontal attenuated total reflection (HATR) stainless steel sample stage. Light-end volatile components will be lost in the evaporation process. 

 

InfraCal Cuvette Holder Analyzer, Model CVH (P/N 405-0003)
This model is recommended for measuring TOG, TPH or FOG levels using the traditional EPA methods 413.2 and 418.1 and Freon-113, spectroscopic hydrocarbon-free grade of perchloroethylene, AK-225, S-316 or other transparent solvent as the extracting solvent. Since the extract is deposited into a 10mm quartz cuvette cell with Teflon stopper, light-end volatile components are retained for measurement. Accurate down to 2 ppm. 

 

InfraCal Card Holder Analyzer, Model CH (P/N 405-0002)
This model uses specially-designed reusable IR sample plates with sapphire windows, having a 100 microliter sample injection capacity. Suited for measuring higher levels of TOG or TPH in water or soil samples extracted with hexane, pentane or Vertrel MCA. Light-end volatile components will be lost in the evaporation process. 

 

InfraCal Analyzers Meeting Industry Needs
InfraCal TOG/TPH Analyzer, Model HATR-T2 Used in WERF Fats, Oil and Grease (FOG) Study
The Water Environment Research Foundation (WERF) sponsored an extensive study to evaluate the performance of different grease interceptor designs and their affect on wastewater discharges. The InfraCal IR Platform Analyzer, Model HATR-T2 was used to measure the concentration levels of fats, oil and grease (FOG) in wastewater leaving the interceptors. The complete study has been released by WERF (www.werf.org) in a report titled “Assessment of Grease Interceptor Performance” (Ducoste, Kenner, Groninger and Holt, 2008)
The InfraCal Cuvette Holder Analyzer, Model CVH The InfraCal Cuvette Holder Analyzer, Model CVH was one of the infrared instruments used to develop ASTM Standard Test Method D 7066--04 --04 
The ASTM has approved a new standard test method for measuring total oil and grease (TOG) and total petroleum hydrocarbon (TPH) concentration levels in water and soil. ASTM Method D 7066-04 covers the use of dimer/trimer of chlorotrifluoroethylene (s-316) for analysis of recoverable oil and grease and nonpolar material by infrared determination.
This ASTM method approves S-316 as a replacement solvent for Freon for this analysis and replaces ASTM Method D 3921. With the publication of this method, the ASTM officially approved the use of S-316 as a replacement solvent for Freon when measuring TOG and TPH in water and soil.
The extraction procedure detailed in the new ASTM method is basically the same as outlined in US EPA Methods 413.2 and 418.1, but using S-316 as the extracting solvent instead of Freon. ASTM Method D 7066-04 is idea for measuring volatile constituents.
Model CVH Allows On--Site Analysis to Meet New European Regulations
New European regulations governing the measurement of oil-in-water, a 15% hydrocarbon emission reduction target and Hydrocarbon Trading have all heightened the importance of monitoring oil concentration levels in water. The Oil Pollution Prevention and Control (OPPC) Regulations prohibit the discharge of oil into the seaunless it is in accordance with the permit issued to cover the discharge. In support of OPPC, the UK DTI (Department of Trade and Industry) has provided industry with technical guidance on suitable methods for the sampling and analysis of produced water and other samples containing hydrocarbons. An OSPAR modified version of the ISO 9377-2 GC-FID method (Gas Chromatography and Flame Ionization Detection) has been adopted as the reference method effective January 1, 2007. Since GCFID analysis is generally done in a laboratory and not onsite, the DTI guidelines permit the use of other methods that have a crude oil specific validation against the GC-FID. Many operators have chosen the InfraCal TOG/TPH Analyzer for use at their facilities because it is a portable device providing faster results at a much lower cost. 

 

Product Specifications

Dimensions: 6.5 x 6.5 x 5” (165 x 165 x 127 mm)
Weight: 4.5 lbs. (2.0 kg)
Display: 4 digit, 7-segment red LED, 5/8 in. character height

Power Requirements:

  • Voltage - 12 V dc, +2% max.
  • Power - 7.5 watts max., 5 watts typical
  • Input - Switchcraft 760 plug or equivalent, center positive

Suggested Power Sources:

  • Wall supply; AC/DC converter type (supplied as standard)
  • 12 volt auto battery adapter connector
  • Portable 12 volt battery pack

Measurement Range: 2 to 5000+ ppm (Dependant on InfraCal Model)
Operating Temperature Range: 40°F (4°C) to 110°F (45°C)

User Selected Calibration:

  • Zero balance adjustment
  • Up to 20 point curve fitting calibration

Communications Port: RS232 port, 9-pin D-Sub female, PC compatible, for upload to PC, datalogger or printer

 