Tubular furnace in an area 900 ° C, 1000 ° C and 1200 ° C Material Testing

Tubular furnace in an area 900 ° C, 1000 ° C and 1200 ° C

Product Code: Models 900 ° C, 1000 ° C and 1200 ° C


Need more Info? (514) 354-2511 - 1-800-463-4363

FEATURES
  • Eleven standard size wire-wound tube furnaces.
  • Maximum operating temperatures of 900ºC, 1000ºC & 1200ºC.
  • Rapid heat-up to operating temperature.
  • Select from multiple work tube diameters ranging from 5/8" (15 mm) to 4" (100 mm).
  • Seven heated length choices from 5" (130 mm) to 36" (900 mm) provide for multiple working and uniformity requirements.
  • Integral ceramic work tube with external element winding creates furnace chamber.
  • Integral work tube is convenient and suitable for processing in air.
  • Easily slide a separate tube through work tube for processing under atmosphere.
  • Furnace incorporates low thermal mass ceramic fiber insulation for improved response times.
  • Furnaces are normally mounted horizontally on a support base or cabinet.
  • Thermocouple is located in a protection tube between the chamber work tube and heating element.
  • All wire-wound tube furnaces can be provided for optional vertical operation.
  • Safety outer mesh guard provides low temperature external surface.
  • Furnace controls are built into the supporting base or cabinet, providing convenient observation and access to all power and temperature controls.
  • Choice of Model 301 control or programmers.
  • Choice of multiple accessories and options, including process tubes, gas tight end seals, insulating plugs, radiation shields, etc.

 

900°C, 1000°C & 1200°C Single Zone Wire-Wound Tube Furnaces
Furnace Model Max. Temp (°C) Work Tube Inside Dia. Inches (mm) Heated Length Inches (mm) Overall Furnace Length Inches (mm) TC Type Max Power (Kw) Furnace Voltage External Dimensions Inches Shipping Weight (lb.)
Height Lenght Depth
MTF 9/15/130 900 0.6 (15) 5.00 (130) 7.00 (180) K .25 120 7.00 (180) 7.00 (180) 3.50 (90) 14
MTF 10/15/130 1000 0.6 (15) 5.00 (130) 6.00 (150) K .40 120 10.50 (265) 6.00 (150) 6.75 (175) 22
• MTF 10/25/130 1000 1.00 (25) 5.00 (130) 6.00 (150) K .40 120 10.50 (265) 6.00 (150) 6.75 (175) 22
MTF 12/25/250 1200 1.00 (25) 10.00 (250) 12.00 (300) N .70 120 14.75 (375) 14.50 (370) 14.75 (375) 38
• MTF 12/38/250 1200 1.50 (38) 10.00 (250) 12.00 (300) N 1.0 120 14.75 (375) 17.75 (450) 14.75 (375) 38
MTF 12/25/400 1200 1.00 (25) 15.75 (400) 17.75 (450) N 1.0 120 17.00 (430) 14.50 (370) 14.75 (375) 44
• MTF 12/38/400 1200 1.50 (38) 15.75 (400) 17.75 (450) N 1.5 120 17.00 (430) 17.75 (450) 14.75 (375) 44
MTF 12/38/850 1200 1.50 (38) 33.50 (850) 35.50 (900) N 2.8 208/240 17.00 (430) 35.50 (900) 14.75 (375) 104
• CTF 12/65/550 1200 2.50 (65) 21.50 (550) 24.50 (625) N 2.0 208/240 20.75 (525) 24.50 (625) 14.25 (360) 80
CTF 12/75/700 1200 3.00 (75) 27.50 (700) 30.50 (775) N 3.0 208/240 20.75 (525) 30.50 (775) 14.25 (360) 84
CTF 12/100/900 1200 4.00 (100) 35.50 (900) 38.50 (975) N 4.5 208/240 20.75 (525) 38.50 (975) 14.25 (360) 104
• Stock Products | Note: Stock furnaces are provided with 301 control. | Specify voltage at time of order. | Continuous operating temperature is 100°C below maximum temperature. Furnaces operate on single phase voltage

 

 

 

 

 