Enregistreur de données pour thermocouple style USB

Enregistreur de données pour thermocouple style USB

Modèle :EL-USB-TC


Besoin de plus d'informations? (514) 354-2511 - 1-800-463-4363

This data logger measures and stores over 32,000 temperature readings from either a K, J or T type thermocouple . A thermocouple is attached via the thermocouple socket at the base of the unit. The user can easily set up the thermocouple type, logging rate, start-time, logging mode, and download the stored data by plugging the module straight into a PC's USB port and running the purpose designed software under Windows 2000, XPor Vista. Data can then be graphed, printed and exported to other applications. The data logger is supplied complete with a long-life lithium battery, which will last for approximately 6 months.
 

Features
  • -200 to +1300°C Measurement Range (K-type)
  • -130 to +900°C Measurement Range (J-type)
  • -200 to +350°C Measurement Range (T-type)
  • USB Interface for Set-up and Data Download
  • 2 User-Programmable Alarm Thresholds
  • Bright Red and Green LED Indication
  • Replaceable Internal Lithium Battery
  • Supplied with basic K-type thermocouple rated from 0 to 110°C (32 to 230°F)

 

Windows Control Software
Easy to install and use, the control software runs under Windows 2000, XP (Home and Professional Editions) and Vista (32-bit). It allows the user to set up and download any EL-USB-TC. The latest version of the control software may be downloaded from www.lascarelectronics.com.

 

Data Logger Set-ups
  • Logger Name
  • °C, °F
  • Logging Rate (1s, 10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)
  • High and Low Alarms
  • Start Date and Start Time
  • Range of logging modes available

 

Specifications

 

LED Flashing Modes
EL-USB-TCfeatures2LEDs.
  • The first LEDflashes orange (O) to indicate a problemcondition onthe EL-USB-TC.
  • The second LED flash es green or red (G/R) to indicate alarm level status.

It will flash red when the logged temperature has exceed eda Lowor High alarm level. By default latching is enabled, so the red LED will continue to flash even after the logged temperature has returned to normal. The red LED will effectively have latched into its alarm condition. This feature ensures that the user is notified that an alarm level has been exceeded, without the need to download the data from the logger.

Latching can be turned off via the control software. The red LED will then no longer continue to flash after the
logged temperature has return ed to normal. Instead, the green LED will flash.
The LEDs on the EL-USB-TC will flash in one of the following ways:

 

DIMENSIONS All dimensions in mm (inches)

 

Battery Replacement
Werecommend that youreplacethe battery every 6 months, or priortologging critical data. The EL-USB-TC does not lose its stored readings when the battery is flat or when the battery is replaced; the data logging process will however be stopped and cannot be re-started until the battery has been replaced and the logged datahas beendownloaded toPC. Check with your supplier that the battery you are ordering is 'press fit' and is not fitted with solder tags. Before replacing the battery, remove theEL-USB-TC fromthePC.

 