Enregistreur température a 4 canaux avec affichage graphique de TandD Biotechnologie

Enregistreur température a 4 canaux avec affichage graphique de TandD

Modèle :MCR-4TC


Besoin de plus d'informations? (514) 354-2511 - 1-800-463-4363

The MCR-4TC Multichannel Recorders are 4-channel battery operated data loggers.


Up to 4 units can be coupled together, making it possible to simultaneously measure and record up to 16 channels.
The MCR-4TC is designed to measure a wide range of temperatures using a thermocouple sensor (compatible types: K, J, T, S and R).

 

Application Examples
  • Record temperatures in pipes and ducts
  • Record boiler temperatures
  • Record temperatures in cooking equipment such as pans, fryers, and ovens
  • For temperature management of refrigerated and frozen goods
  • For temperature testing of the electronic substrates

 

Specifications
Measurement Channels Temperature 4 ch
Compatible Sensors Thermocouple: Type K, J, T, S, R
Measurement Units °C, °F
Measurement Range Type K : -270 to 1370 °C
Type J : -210 to 1200 °C
Type T : -270 to 400 °C
Type S : -50 to 1760 °C
Type R : -50 to 1760 °C
Input Impedance Approx. 1 MΩ
Accuracy (*1)   Thermocouple Measurement
Type K, J, T: ±(0.5 °C + 0.3 % reading) [ at -100 °C or above ]
Type S, R : ±(1.5 °C + 0.3 % reading) [ at 100 °C or above ]
(Individual sensor inaccuracies not included.)
Cold Junction Compensation
±0.5 °C [ in operating environent of 10 to 40 °C ]
±0.8 °C [ in other operating environment ] 
Auto : According to the range in use
Measurement Resolution 0.1 °C
Recording Method Instantaneous or Average
Recording Interval  100, 200, 500 ms.
1, 2, 5, 10, 15, 20, 30 sec.
1, 2, 5, 10, 15, 20, 30, 60 min.
Logging Capacity (*2)   When recording 1 channel : up to 960,000 readings/ch
When recording 2 channels : up to 480,000 readings/ch
When recording 3 channels : up to 320,000 readings/ch
When recording 4 channels : up to 240,000 readings/ch
Recording Mode Endless ( Overwrite oldest data in the current recording session when capacity is full ) 
One Time ( Stop recording when capacity is full )
Recording Start Method  Immediate Start or Programmed Start ( by individual logger or by group )
Group Recording  Up to 4 units ( 16 channels ) can be recorded simultaneously. ( Coupling of MCR-4TC and MCR-4V is possible. (*3) )
LCD Display Items Measurements, Trend Graph, Battery Level, etc.
Communication Interfaces USB Communication
Communication Time Download Times for Full Data via USB
- From the master or a single unit: Approx. 1 minute 30 seconds
- From a slave unit: Approx. 4 minutes 30 seconds
External Memory  SD Memory Card, SDHC Memory Card (*4) ( For Manual or Automatic Data Export )
Power AA Alkaline Battery x 2 ( AA Ni-MH batteries may also be used ), USB Power ( 5 V 250 mA )
Battery Life (*5)  For 4-Channel Measurement (AA Alkaline Battery)
  Rec.Method
Instanteneous Value  Average Value
Rec Interval 100 ms  Approx. 5 days Approx. 5 days
500 ms  Approx. 7 days Approx. 7 days
1 sec Approx. 21 days Approx. 7 days
5 sec or longer Approx. 60 days Approx. 21 days

(Battery life for 1-channel measurement is about 1.4 times longer than that of 4-channel measurement.)

Input Terminal Screwless Terminals
< Compatible Wires >
Single Wire : ø 0.32mm to ø 0.65mm ( AWG 28 - 22 )
Twisted Wire : 0.08mm2 to 0.32mm2 ( AWG 28 - 22 ), ø 0.12 mm or more in diameter
Stripping Length : 9 to 10 mm
Isolation CH1, CH2, CH3, CH4, and USB are isolated.
( Battery terminals are not isolated from the CH1-CH4 input terminals. )
CH1-CH4 Maximum Applied Voltage : ±50 V
Electrical Isolation Resistance : 50 MΩ or more ( DC ± 250 V )
Dimensions H 120 mm x W 75 mm x D 32 mm
Weight Approx. 190 g ( including batteries )
Operating Environment Temperature: 0 to 50 °C
Humidity: 90 %RH or less ( no condensation )
Accessories AA Alkaline Battery x 2, USB Communication Cable (US-15C), Software (CD-ROM), Card Slot Cover, User's Manual Set (Warranty Included)
*1 : MCR-4TC has superior noise filter, but the measurement may sometimes fluctuate due to strong noise. Especially when the recording interval is set to 200 ms or less, the filtering becomes weaker and hence the fluctuation may become greater.
*2 : If the logging capacity is not filled at the end of one recording session, the logger can record up to 30 times.
*3 : Group Recording may not be started depending on the recording or measurement interval specifications of the connected Master unit.
*4 : Please check the [Functions] for information on memory cards whose operation has been confirmed.
*5 : Battery life varies depending upon multiple factors including ambient temperature, recording interval, number of measurement channels, and frequency of data export to a memory card. All estimates are based on operations carried out with a new battery and are in no way a guarantee of actual battery life

Convertisseur d'unités