Test Method
Heat of combustion is determined in this test method by burning a weighed sample in an oxygen bomb calorimeter under controlled conditions. The heat of combustion is computed from temperature observations before, during and after combustion with proper allowances for thermochemical and heat transfer corrections. Either isothermal or adiabatic calorimeter jackets can be used.

Automatic Calorimeter

The automatic calorimeter is the latest system for determining gross calorific values of liquids and solid fuels. A higher level of automation with extremely simple handling characterizes this device.

In addition to the Isoperibolic measurement procedure, a Dynamic (reduced-time) mode is also available for the user. Different working temperatures can be selected for both procedures based on the temperature of the connected water.
To provide a supply of cooling water, the calorimeter can be connected to a standard thermostat or an appropriate permanently installed water connection, with a connection valve. The unit is equipped with a very convenient operating panel through which operation of the device takes place. The graphical display with active back lighting displays the appropriate status messages. The temporal course of a measurement that has been started and all current parameters of the weighed in sample can be constantly monitored and are arranged to be clearly visible.

Connections for analysis scale, printer, sample rack for identifying and managing samples are already integrated into the basic device. The network connection and the special configuration for data exchange can be implemented at any time with LIMS.

In combination with special halogen-resistant decomposition vessels quantitative decompositions can be performed to determine halogen and sulfur content.


Specifications
  • Conforms to the specifications of:
    • ASTM D240; D4809; D5865; D1989; D5468; E711; ISO 1928;
    • DIN 51900; BS1016
  • Measurement range: 40,000 J
  • Measuring mode: Isoperibolic 25°C; Isoperibolic 30°C; Dynamic 25°C; Dynamic 30°C
  • Isoperibolic Measuring Time: Approximately 22 min
  • Dynamic Measuring Time: Approximately 7 min
  • Oxygen Operating Pressure: 30 bar
  • Cooling Medium: Water via line, flow through quantity 60 + 10 liters / hour
  • Water Operating Pressure: 1 – 1.5 bar max. b10 bar
  • Interfaces: Serial (RS232); Parallel; Keyboard; Sample rack; External monitor

Dimensions
  • L x W x H, in.(cm)
  • 151⁄2 x 10 x 41⁄2 (391⁄2 x 251⁄2 x 111⁄2)
  • Net Weight: 23 lbs (101⁄2kg)

Shipping Information
  • L x W x H, in.(cm)
  • 171⁄2 x 173⁄4 x 193⁄4 (440 x 450 x 500)
  • Net Weight: 66 lbs (30 kg)

Ordering Information
Catalog No.  
K88800
Automatic Calorimeter, 115V 60Hz
K88890
Automatic Calorimeter, 220V 50Hz
  Accessories
K88800-1

Cooling water supply unit, 115V 60Hz

K88890-1
Cooling water supply unit, 220V 50Hz
K88800-2
Pressure Gauge, Oxygen
To reduce the pressure of the oxygen cylinder to 30 bar
K88800-3
Standard Decomposition Vessel
K88800-4
Decomposition Vessel, Halogen Resistant
For quantitative decomposition determine halogen and sulfur content
K88800-5
Connection valve
Required for permanently installed water connection

 

Convertisseur d'unités